Bulk etch rate measurements and calibrations of plastic nuclear track detectors
/ Authors
S. Balestra, M. Cozzi, G. Giacomelli, R. Giacomelli, M. Giorgini, Ashavani Kumar, Ashavani Kumar, G. Mandrioli, S. Manzoor, S. Manzoor
and 10 more authors
A. Margiotta, E. Medinaceli, L. Patrizii, V. Popa, I. Qureshi, M. A. Rana, G. Sirri, M. Spurio, V. Togo, C. Valieri
/ Abstract
Abstract New calibrations of CR39 and Makrofol nuclear track detectors have been obtained using 158 A GeV Pb82+ and In49+ions; a new method for the bulk etch rate determination, using both cone height and base diameter measurements, was developed. The CR39 charge resolution based on the etch-pit base area measurements is adequate to identify nuclear fragments in the interval 7 ⩽ Z/β ⩽ 49. For CR39 the detection threshold is at REL ∼50 MeV cm2 g−1, corresponding to a nuclear fragment with Z/β ∼ 7. Base cone area distributions for Makrofol foils exposed to Pb82+ ions have shown for the first time all peaks due to nuclear fragments with Z > 50; the distribution of the etched cone heights shows well separated individual peaks for Z/β ∼ 78 to 83 (charge pick-up). The Makrofol detection threshold is at REL ∼2700 MeV cm2 g−1, corresponding to a nuclear fragment with Z/β = 50.
Journal: Nuclear Instruments & Methods in Physics Research Section B-beam Interactions With Materials and Atoms