Energy calibration of the JLab bremsstrahlung tagging system
/ Authors
S. Stepanyan, S. Boyarinov, H. Egiyan, H. Egiyan, L. Guo, D. Dale, M. Gabrielyan, L. Gan, A. Gasparian, A. Glamazdin
and 4 more authors
/ Abstract
In this report, we present the energy calibration of the Hall B bremsstrahlung tagging system at the Thomas Jefferson National Accelerator Facility. The calibration was performed using a magnetic pair spectrometer. The tagged photon energy spectrum was measured in coincidence with e + e pairs as a function of the pair spectrometer magnetic field. Taking advantage of the internal linearity of the pair spectrometer, the energy of the tagging system was calibrated at the level of ±0.1%E. The absolute energy scale was determined using the e + e rate measurements close to the end-point of the photon spectrum. The energy variations across the full tagging range were found to be < 3 MeV.
Journal: Nuclear Instruments & Methods in Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment