Flexible Piezoresistive Yarn Sensor for Human Physiological Signal Measurement
Rizal Maulana, Ádám Rák, Sándor Földi, György Cserey
Abstract
Continuous monitoring of physiological signals is essential for the early detection of health problems. A measurement system that ensures high sensitivity, accuracy, and user comfort is needed. In this study, we designed and optimized a flexible piezoresistive yarn (FPY) sensor to achieve a high sensitivity and wide working range for detecting physiological signals. The representative sensor design was constructed by applying an FPY bonding pattern, utilizing tightly arranged triangular patterns and using minimal FPY. The prototype sensor operates in two measurement modes, strain and pressure, and was evaluated for measuring neck motion, finger bending, respiratory signals, and arterial blood pressure (ABP) waveforms. A qualitative evaluation, performed by comparing the characteristics of the measurement results of each physiological signal with those from related studies, indicates a high similarity in its morphological characteristics. Then, a quantitative evaluation through baseline drift analysis demonstrates that the FPY sensor displays high measurement stability. The ABP waveform measurement shows the most stable baseline, with a mean absolute error (MAE) of $0.0051 \pm 0.0029$ in terms of baseline drift, using normalized values from 0 to 1. Based on our results, the prototype sensor can be used as an innovative solution for physiological signal monitoring and can be further enhanced for personalized healthcare and sports applications.