Neutron spectrometer for fast nuclear reactors
/ Authors
M. Osipenko, M. Ripani, G. Ricco, B. Caiffi, F. Pompili, M. Pillon, M. Angelone, G. Verona-Rinati, R. Cardarelli, G. Mila
and 1 more author
/ Abstract
Abstract In this paper we describe the development and first tests of a neutron spectrometer designed for high flux environments, such as the ones found in fast nuclear reactors. The spectrometer is based on the conversion of neutrons impinging on 6 Li into α and t whose total energy comprises the initial neutron energy and the reaction Q -value. The 6 LiF layer is sandwiched between two CVD diamond detectors, which measure the two reaction products in coincidence. The spectrometer was calibrated at two neutron energies in well known thermal and 3 MeV neutron fluxes. The measured neutron detection efficiency varies from 4.2 × 10 − 4 to 3.5 × 10 − 8 for thermal and 3 MeV neutrons, respectively. These values are in agreement with Geant4 simulations and close to simple estimates based on the knowledge of the 6 Li(n,α) t cross-section. The energy resolution of the spectrometer was found to be better than 100 keV when using 2.5 or 5 m cables between the detector and the preamplifiers.
Journal: Nuclear Instruments & Methods in Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment