Bent crystal spectrometer for both frequency and wavenumber resolved x-ray scattering at a seeded free-electron laser.
/ Authors
U. Zastrau, L. Fletcher, E. Förster, E. Galtier, E. Gamboa, S. Glenzer, P. Heimann, H. Marschner, B. Nagler, A. Schropp
and 2 more authors
/ Abstract
We present a cylindrically curved GaAs x-ray spectrometer with energy resolution ΔE/E = 1.1 × 10(-4) and wave-number resolution of Δk/k = 3 × 10(-3), allowing plasmon scattering at the resolution limits of the Linac Coherent Light Source (LCLS) x-ray free-electron laser. It spans scattering wavenumbers of 3.6 to 5.2/Å in 100 separate bins, with only 0.34% wavenumber blurring. The dispersion of 0.418 eV/13.5 μm agrees with predictions within 1.3%. The reflection homogeneity over the entire wavenumber range was measured and used to normalize the amplitude of scattering spectra. The proposed spectrometer is superior to a mosaic highly annealed pyrolytic graphite spectrometer when the energy resolution needs to be comparable to the LCLS seeded bandwidth of 1 eV and a significant range of wavenumbers must be covered in one exposure.
Journal: The Review of scientific instruments
DOI: 10.1063/1.4894821